Applications such as pulp and paper or dye manufacturing and industrial wastewater treatment may expose pH probes to harsh conditions and chemicals. A differential pH probe utilizes a special measurement technique, which extends the working lifetime of the sensor in challenging applications. Replacing the sensor reference (salt bridge) and buffered electrolyte solution can extend working lifetime, which saves timely, costly replacements. Our SD7420CD features differential sensor technology with a direct 4-20mA output, so it can be integrated into control systems without the use of a transmitter or controller.
CA Residents: Prop 65 WARNING(S)
For optimal shelf life, store sensors at 15-30°C.
Measurement Range: 0-14 pH (low NA+ ion error)
Temperature Range: 0-100⁰ C
Pressure Range: 0 – 100 PSIG @ 40⁰ C (PVC fitting); 0 – 100 PSIG @ 60⁰ C (CPVC fitting); 0 – 100 PSIG @ 100⁰ C (stainless steel fitting);
Body Construction: Ryton Polyphenylene Sulfide (PPS)
Response Time: 95% in 1 second
Isopotential: 7.00 pH
Offset: 0.2%
Span: > 97%
Temperature Compensation: 300Ω NTC
Connector: Tinned Leads
Cable Length: 20 feet (6 meters)
Power Supply (not included): 12- 30 V DC ( >24V DC recommended)
Installation: 1 1/2″ NPT
Warranty: Two year limited warranty
| Weight | 1.90 lbs |
|---|---|
| Dimensions | 10 × 4 × 3 in |
Applications such as pulp and paper or dye manufacturing and industrial wastewater treatment may expose pH probes to harsh conditions and chemicals. A differential pH probe utilizes a special measurement technique, which extends the working lifetime of the sensor in challenging applications. Replacing the sensor reference (salt bridge) and buffered electrolyte solution can extend working lifetime, which saves timely, costly replacements. Our SD7420CD features differential sensor technology with a direct 4-20mA output, so it can be integrated into control systems without the use of a transmitter or controller.