TX3100 - pH/ORP Transmitter/Controller, Intelligent

TX3100 - pH/ORP Transmitter/Controller, Intelligent

TX3100
NOW Available! Our TX3100 is an intelligent pH and ORP controller/transmitter for reading pH and ORP sensors and programmable process control. This controller replaces the TX3000 1/2 DIN pH/ORP controller and features an updated user interface with user friendly full text and graphical display. This makes operation, calibration, and maintenance easier for operators in treatment plants. The TX3100 also features a solution ground input for accurate readings in applications with ground loop issues. NEW graphical display shows sensor data and aids in the calibration process 2 programmable relays (Hi/Lo, Hi/Hi, or Lo/Lo) Solution ground input eliminates measurement errors caused by reference ground potentials Compatible with all Sensorex sensors with tinned leads and PT1000 RTD temp comp
Agotado
0 Kč
Impuestos incluidos
Cantidad

Fuera de stock

Our TX3100 is an intelligent pH and ORP controller/transmitter for reading pH and ORP sensors and programmable process control. This controller replaces the TX3000 1/2 DIN pH/ORP controller and features an updated user interface with user friendly full text and graphical display. This makes operation, calibration, and maintenance easier for operators in treatment plants. The TX3100 also features a solution ground input for accurate readings in applications with ground loop issues.

  • Reliable pH/ORP monitoring and process control: Proven technology for pH/ORP monitoring and process control with sample averaging for increased stability
  • Simple sensor calibration: NEW graphical display shows sensor data and aids in the calibration process
  • High-accuracy measurements: Solution ground input eliminates measurement errors caused by reference ground potentials
  • Compatible with a wide variety of sensors: For use with all Sensorex combination-style pH/ORP sensors
  • One year limited warranty: We stand behind all our products with the Sensorex Guarantee

Note: Transmitter is only compatible with sensors with tinned leads.

CA ResidentsProp 65 WARNING(S)



  • Measurement Range pH: -2.00 to 16.00pH, 0.01 pH resolution

    Measurement Range ORP/mV: -1999mV to 1999mV, 1mV resolution, +/- 1mV accuracy

    Temperature Range: -30 °C to 130 °C , 0.1 °C resolution, 0.2 °C accuracy

    Calibration Mode: Any Tech & NIST or custom buffers up to 3 points

    Enclosure: 1/2 DIN NEMA 4X IP65

    HWD: 144 mm (3.78″) x 144 mm (3.78″) x 115mm (5.20″)

    Panel Cut-Out: 138mm (3.66″) x 138mm (3.66″)

    Display: Full graphical and text display

    Analog Output 1: 0/4 -20mA, corresponding to pH or ORP output. Max load 500 Ω.

    Analog Output 2: 0/4 -20mA, corresponding to temperature output. Max load 500 Ω

    Voltage Output: DC +/- 12V, 1W max

    Alarm/Control Relays: Contact: 240V AC, 5A max Activate: 2 setpoints Hi/Low, On/Off, Hi/Hi, Low/Low

    Wash Setting: Contact: 240V AC, 30V DC 5.0A max On: 0-9999 seconds /Off: 0-9999 hours

    Mounting Options: Panel or wall mounting

    Ambient Operating Temperature: 0° C to 50° C

    Ambient Humidity: 0 to 95%, non-condensing

    Temperature Compensation: Pt1000Ω RTD with automatic compensation or manual

    Power Requirement: 100-240V AC +/- 10%, 9W max., 50/60 Hz

    Warranty: One year limited warranty

    Weight: Approx. 1.75lb (0.80kg)

    Certifications: CE



Weight2.76 lbs
Dimensions11 × 10.5 × 6.5 in

NOW Available!

Our TX3100 is an intelligent pH and ORP controller/transmitter for reading pH and ORP sensors and programmable process control. This controller replaces the TX3000 1/2 DIN pH/ORP controller and features an updated user interface with user friendly full text and graphical display. This makes operation, calibration, and maintenance easier for operators in treatment plants. The TX3100 also features a solution ground input for accurate readings in applications with ground loop issues.

  • NEW graphical display shows sensor data and aids in the calibration process
  • 2 programmable relays (Hi/Lo, Hi/Hi, or Lo/Lo)
  • Solution ground input eliminates measurement errors caused by reference ground potentials
  • Compatible with all Sensorex sensors with tinned leads and PT1000 RTD temp comp
: TX3100
Agotado

Ficha técnica

Enclosure
1/2 DIN NEMA 4X IP65
Ambient Humidity
0 to 95%, non-condensing
Ambient Operating Temperature
0° C to 50° C
Certifications
CE
Measurement Range pH
-2.00 to 16.00pH, 0.01 pH resolution
Measurement Range ORP/mV
-1999mV to 1999mV, 1mV resolution, +/- 1mV accuracy
Calibration Mode
Any Tech & NIST or custom buffers up to 3 points
HWD
144 mm (3.78″) x 144 mm (3.78″) x 115mm (5.20″)
Panel Cut-Out
138mm (3.66″) x 138mm (3.66″)
Analog Output 1
0/4 -20mA, corresponding to pH or ORP output. Max load 500 Ω.
Display
Full graphical and text display
Analog Output 2
0/4 -20mA, corresponding to temperature output. Max load 500 Ω
Contact
240V AC, 5
240V AC, 30V DC 5.0
Dimensions
11 x 10.5 x 6.5 in
Voltage Output
DC +/- 12V, 1W max
Applications
Controllers & Transmitters, ORP, pH
A max Activate
2 setpoints Hi/Low, On/Off, Hi/Hi, Low/Low
Warranty
One year limited warranty
Mounting Options
Panel or wall mounting
Power Requirement
100-240V AC +/- 10%, 9W max., 50/60 Hz
0-9999 seconds /Off
0-9999 hours
Temperature Range
-30 °C to 130 °C , 0.1 °C resolution, 0.2 °C accuracy
Temperature Compensation
Pt1000Ω RTD with automatic compensation or manual
Peso
2.76 lbs
Approx. 1.75lb (0.80kg)